How often should a CMM be calibrated?
Most inspection departments run an annual reverification, with interim checks using ball plates or ring gauges between formal intervals. The right interval depends on usage, environment and customer requirement — the laboratory can recommend one based on the machine's history.
What is the difference between calibration and verification?
Calibration establishes the relationship between a machine's readings and reference values, with a stated uncertainty. Performance verification checks whether the machine still meets a stated specification. Most CMM work ordered as calibration is actually verification.
What does ISO 10360-2 actually test?
Length measurement error, designated E0 and E150, is measured at five calibrated lengths, three repeats each, across seven positions and orientations — 105 measurements compared against the machine's maximum permissible error.
Can it be done on site, and how long is the machine down?
Bridge and gantry machines are typically tested on-site. Time on site and scheduling around shutdown windows are confirmed at quote, matched to the machine's size and the test scope required.
What temperature does the room need to be?
ISO 1 fixes 20°C as the dimensional reference temperature. For an acceptance test the manufacturer's specified environment applies; for reverification the user may specify conditions, which is why they are recorded on the certificate.
What happens if the machine fails as-found?
As-found deviations are recorded before adjustment. That record supports an impact assessment back to the last passing calibration — which parts were accepted, and whether re-inspection follows.
Is there CMM calibration near me?
On-site service runs regularly through Louisville, Lexington, Bowling Green, Owensboro, Elizabethtown, Columbus and Evansville and their surrounding towns. Machines outside that radius are typically scheduled as ship-in work under the same certificate format.
What is a CMC and why does it matter to my tolerances?
Calibration and measurement capability is the smallest uncertainty a laboratory can claim within its scope. When a tolerance runs tighter than the available CMC, a 4:1 test uncertainty ratio may not hold, and that should be flagged before the work is booked.
Will this satisfy my IATF or AS9100 auditor?
Auditors check the laboratory's scope of accreditation against the measurement, range and uncertainty actually needed, not the certificate alone. Scope documents are available on request to support that check.
What is an error map, and can it be updated on my controller?
The volumetric compensation map corrects for geometric error across the measuring volume. Many controllers allow an update during geometric adjustment; some are OEM-locked and require the original manufacturer's software.