Primary Service
CMM Calibration and Repair
Request a quote for CMM calibration or repair.
Request a quotation

Calibration versus verification
Calibration establishes the relationship between a machine's indications and a reference value, with a stated uncertainty. What is usually sold as CMM calibration is a performance verification against ISO 10360-2 or ASME B89.4.1, often combined with adjustment.
Both routes are offered. Older machines are frequently held to B89.4.1 by customer preference; newer installations verify against ISO 10360-2 directly.
What a visit covers
Scales, geometry, probing and length measurement error are checked in sequence, with results recorded before any adjustment is made.
- 01
Scale and reader head
X, Y and Z scales are inspected and cleaned; reader head output is checked and adjusted.
- 02
Geometry
Straightness, linearity and squareness are checked in the XY, ZX and ZY planes.
- 03
Length measurement error
E0 and E150 tested per ISO 10360-2, 105 measurements against stated MPE.
- 04
Probing error
PFTU tested per ISO 10360-5 on a dedicated calibrated test sphere.
- 05
Repair scope
Probe heads, air bearings, drives, counterbalance and controller retrofit after crash or relocation.
Sequence of a calibration visit
- 01
Thermal soak
Artifacts and machine are allowed to stabilize to room temperature before testing begins.
- 02
As-found testing
Geometry, length measurement error and probing error are recorded before adjustment.
- 03
Adjustment
Geometric compensation and error map updates are applied where the controller supports them.
- 04
As-left verification
Tests are repeated and results issued on a certificate with stated uncertainty.

Why Louisville CMM Calibration and Repair
- 01
As-found recorded first
Out-of-tolerance findings are documented before adjustment, supporting a reverse-traceability review.
- 02
Both standards offered
Verification is available against ISO 10360-2 or ASME B89.4.1 depending on the certificate the machine was built to.
- 03
Repair alongside calibration
Air bearings, drives and probe heads are serviced in the same visit where practical.
Coverage across Kentucky and the Greater Midwest
On-site and ship-in service is arranged across Kentucky and the Greater Midwest. Travel radius and scheduling are confirmed at quote.
Other services
Discuss CMM Calibration Requirements
Send the machine make, model and axis travel to start a quote.